Dental Journal of Advance Studies

Register      Login

VOLUME 8 , ISSUE 1 ( January-April, 2020 ) > List of Articles

Original Article

To Evaluate the Influence of Different Implant Thread Designs on Stress Distribution of Osseointegrated Implant: A Three-Dimensional Finite-Element Analysis Study–An In Vitro Study

Chhavi Sharma, Anupreet Kaur Chawla

Keywords : dental implant, FEA study, implant thread, stress analysis, von mises stress

Citation Information : Sharma C, Chawla AK. To Evaluate the Influence of Different Implant Thread Designs on Stress Distribution of Osseointegrated Implant: A Three-Dimensional Finite-Element Analysis Study–An In Vitro Study. 2020; 8 (1):9-16.

DOI: 10.1055/s-0040-1709218

License: CC BY-NC-ND 4.0

Published Online: 02-04-2020

Copyright Statement:  © 2020 Bhojia Dental College and Hospital affiliated to Himachal Pradesh University


Abstract

Introduction Dental implants are common treatment modality for tooth loss which leads to unaesthetic appearance and may also cause deterioration of mastication and speech. The aim of implant therapy in dentistry is to restore tissue contour, function, comfort, aesthetic, and speech. Dental implant role is to transfer the mechanical force created during chewing to the supporting osseous tissues within the mandible and maxilla. The importance of biomechanical factors such as the bone-implant interface, implant thread design, the length and diameter of implants, type of loading, the quality and quantity of surrounding bone have been strained by various authors. The selection of implant thread design plays an important role in the outcome of the treatment. This study was done to evaluate the influence of different thread designs on stress distribution of osseointegrated implant using three-dimensional (3D) finite-element analysis. Materials and Methods Three implants with different thread designs, namely V-thread, buttress, and reverse buttress thread designs were considered and dimensions were standardized. The site considered was the mandibular molar region with cortical and cancellous bone assuming to be isotropic and homogeneous. The implant modeling was done with the ANSYS 18.1 software. Axial load (100N) and buccolingual load (50N) were applied. The stresses were calculated as Von Mises stress criterion. Results Minimum von mises Stress concentration was seen for tapered implant body with reverse buttress thread design under axial load 100N and tapered implant body with V-thread under buccolingual load of 50N at cortical bone which signifies bone preservation. Stress levels were observed maximum at implant and minimum at the cancellous bone. Conclusion Hence, within the limitations of this study the results obtained can be applied clinically for appropriate selection of implant thread design for a predictable success of implant therapy.


PDF Share
  1. The effect of thread pattern upon implant osseointegration. Clin Oral Implants Res 2010;21(2):129–136
  2. Tapered implants in dentistry: revitalizing concepts with technology: a review. Adv Dent Res 2016;28(1):4–9
  3. Finite element analysis of four thread-form configurations in a stepped screw implant. J Oral Rehabil 2004;31(3):233–239
  4. References and techniques used in alveolar bone classification. J Interdiscipl Med Dent Sci 2015;3:2
  5. Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent 2001;85(6):585–598
  6. The influence of thread geometry on implant osseointegration under immediate loading: a literature review. J Adv Prosthodont 2014;6(6):547–554
  7. Influence of three different implant thread designs on stress distribution: A three-dimensional finite element analysis. J Indian Prosthodont Soc 2016;16(4):70–75
  8. Influence of implant length, diameter, and geometry on stress distribution: a finite element analysis. Int J Periodontics Restorative Dent 2010;30(1):89–95
  9. A three-dimensional finite element study on the biomechanical simulation of various structured dental implants and their surrounding bone tissues. Int J Dent 2016;2016:4867402
  10. The influence of implant diameter and length on stress distribution of osseointegrated implants related to crestal bone geometry: a three-dimensional finite element analysis. J Prosthet Dent 2008;100(6):422–431
  11. Radiographic evaluation of marginal bone levels around dental implants with different designs after 1 year. Acta Odontol Scand 2013;71(1):92–95
PDF Share